Technical Specification; Bottom View - Varian Turbo-V 1000 ICE Mode D'emploi

Masquer les pouces Voir aussi pour Turbo-V 1000 ICE:
Table des Matières

Publicité

TECHNICAL SPECIFICATION

Pumping speed
N
: 950 l/s
2
He: 870 l/s
H
: 900 l/s
2
Compression ratio
N
: >1 x 10
2
He: 5 x 10
H
: 2 x 10
2
Base pressure*
minimum with recommended
mechanical forepump:
< 1 x 10
minimum with recommended
diaphragm forepump:
2 x 10
Inlet flange
ISO 200; CFF 10"
Foreline flange
KF 40 NW
Rotational speed
38000 RPM
Start-up time
< 4 minutes
Recommended
mechanical: Varian DS 402
forepump
diaphragm: Varian MD 60
Operating position
any
Cooling
Water (use water with electrical con-
ductivity ≤ 500 µS/cm)
requirements
10
7
6
-9
-9
mbar (< 1 x 10
Torr)
-8
-8
mbar (1.5 x 10
Torr)
Operating ambient
temperature
Coolant water
Bakeout temperature
Vibration level
(displacement)
Noise level
Input
Lubricant
Storage temperature
Weight kg (lbs)
* According to standard DIN 28 428, the base
pressure is that measured in a leak-free test
dome, 48 hours after the completion of test dome
bake-out, and using the recommended pre-
vacuum pump.
26
TECHNICAL INFORMATION
UPPER VIEW
BOTTOM VIEW
+ 5° C to + 35° C
minimum flow: 200 l/h (0.89 GPM)
temperature: + 10° C to + 20° C
pressure: 3 to 5 bar (45 to 75 Psi)
80° C at inlet flange
(with ISO flange)
120° C at inlet flange
(with CF flange)
< 0.01 µm at inlet flange
≤ 45 dB (A) at 1 meter
56 Vac, three phase, 633 Hz
permanent lubrication
- 20° C to + 70° C
ISO 200: 19 (41.8)
CFF 10": 26 (57.2)
87-900-900-01(C)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9080969-9100

Table des Matières