Typical Layout Diagram; Connection A - High Vacuum Flange - Varian Turbo-V 1000 ICE Mode D'emploi

Masquer les pouces Voir aussi pour Turbo-V 1000 ICE:
Table des Matières

Publicité

TYPICAL LAYOUT DIAGRAM

1.
Turbo-V controller
2.
Vent valve
3.
Vacuum pump shut-off valve (optional)
4.
System vent valve (optional)
5.
Vacuum chamber
6.
Ionisation gauge
7.
Fore-vacuum pump connecting flange
8.
Oil mist eliminator
9.
Fore-vacuum pump with internal one-way
valve
10. Fore-vacuum pump control relay
11. Connection for water cooling
12. Roughing line with valve (optional)
13. Turbopump
14. Flexible connection
15. Solid state flowmeter
16. Gas purge valve (integrated into the pump)

Connection A - HIGH VACUUM FLANGE

To connect the Turbo pump to the ISO inlet flange,
remove the outer ring and position the centering
ring as shown in the figure.
Then fix the two flanges with the clamps or claws
as shown in the figure.
For ConFlat flange connections we recommend
using Varian hardware. Other hardware can be
used if it satisfies to this specifications:
− stainless steel screws
− minimum strength point of 500 N/mm
− length: 63.5 mm (2.5 inches).
To facilitate assembly and dismantling, apply Fel-
pro C-100 high temperature lubricant to the screw
threads protruding from the flange and between
the nuts and flange.
Note that the connections can be made only with
the bolt in the lower side.
Attach the units and tighten each one in turn. Re-
peat the sequential tightening until the flange faces
meet.
Exercise care when tightening nuts and bolts to
avoid creating dents in the envelope as this may
cause the pump rotor to lock.
32
TECHNICAL INFORMATION
2
CAUTION
87-900-900-01(C)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9080969-9100

Table des Matières