FILLING: Multiple-Use Disks
When the syringe is placed in the pool or spa sample, and the plunger
is pulled all the way up, the syringe will hold more than enough
sample water to adequately fi ll one section of the disk. The syringe is
held vertically and the tip is inserted into the fi ll hole on the disk. The
plunger is pressed slowly and smoothly to fi ll the disk.
1. Fill any unused section of the disk using slow, even
pressure. Add the sample water until the Blank
Chamber is full and the Over Flow Chamber is partially
fi lled.
2. Do not overfi ll the disk. If the disk is overfi lled, sample
water will fl ow out of the vent hole in the center of the
disk. The disk is not leaking. Dry the disk and run the
test.
3. Do not under fi ll the disk. If the disk is under fi lled, the
blank chamber will not be full. The reagent chambers
will not fi ll entirely and results will be inaccurate.
4. Do not introduce air bubbles into the disk. The reagent
chambers will not fi ll entirely and results will be
inaccruate. Remove the water sample with the syringe
and refi ll.
5. Wet disks should be dried thoroughly with a lint free
wipe. Handle the disk by the edges. Between tests
store the disk on a fl at surface. Do not agitate a
partially fi lled disk to avoid leaking and contamination.
Partially used disks should be discarded at the end of
the day.
For fi lling and troubleshooting tips go to:
www.lamotte.com/spin
7
Over Flow
Chamber
Blank
Chamber
Correct Fill
Over Flow
Chamber
Blank
Chamber
Over
Filled
(Dry disk before placing in meter)
Blank
Chamber
Under Filled
Air Bubble