FILLING: Single-Use Disks
When the syringe is placed in the pool or spa sample, and the
plunger is pulled all the way up, the syringe will hold more than
enough sample water to adequately fi ll the disk. The syringe is
held vertically and the tip is inserted into the fi ll hole on the disk.
The plunger is pressed slowly and smoothly to fi ll the disk.
1. Fill the disk using slow, even pressure. The
sample water will fi ll the spaces between
the baffl es in a counterclockwise order.
Each space will fi ll from the bottom to the
top. Sample water should be added until the
sample water in the fourth chamber fi lls to the
top of the chamber slightly past the embossed
fi ll line. It is OK to fi ll slightly past the fi ll line.
2. Do not overfi ll the disk. If the disk is overfi lled,
sample water will fl ow out of the overfl ow
hole in the center of the disk. The disk is not
leaking. Dry the disk and run the test.
3. Do not under fi ll the disk. If the disk is under
fi lled, the reagent chambers will not fi ll entirely
and results will be inaccurate.
4. Wet disks should be dried thoroughly with a
lint free wipe. The disk should be handled by
the edges.
5. Disks should be fi lled and used within 10 minutes. They cannot be fi lled ahead of
time.
For fi lling and troubleshooting tips go to:
Correct Fill
(Fill line)
Over
Filled
Under
Filled
www.lamotte.com/spin
6
(Dry disk before placing in meter)