TROUBLESHOOTING WITH THE METER CHECK DISK
Do not attempt to separate the components of the Meter Check Disk (Code 1705). The
Meter Check Disk consists of a disk with a permanently attached cover.
Do not fi ll the Meter Check Disk with water. Water is not used during the meter check
procedure.
Range Check Procedure
The Meter Check Disk is used to verify the performance of a meter. It is not used to
recalibrate the meter. Meters are calibrated at the time of manufacture. The meter is
performing satisfactorily if readings using the Meter Check Disk are within the ranges
provided on the Meter Check Disk pouch. Range specifi cations are specifi c to the disk
identifi ed by the serial number on the pouch. The range specifi cations will vary from disk to
disk. The exact readings from a specifi c disk may vary from meter to meter.
1. Follow the Cleaning procedure on page 22 to clean the light chamber and optic lenses.
2. Tap
select Disk Series.
3. Tap MCD to select Meter Check Disk.
MC
M
4. Tap
to return to the test screen.
5. Remove the Meter Check Disk from the foil pouch. DO NOT remove the black cover
from the disk.
6. Insert the Meter Check Disk. Close the lid.
7. Tap
to start test.
8. Results will be displayed.
9. Compare the results on the display to the values on the chart shown on the pouch.
10. Replace the Meter Check Disk in the foil pouch for storage.
If the results are not within the range shown on the pouch, contact tech service.

Alignment Procedure

The Meter Check Disk is inserted in the meter and Calibration is chosen from the Settings
Menu to perform an angle calibration which checks the alignment of the hub and disk. The
results are analyzed and reported as Pass or Fail. If the measurements pass, the settings
will be saved. If the analysis fails, contact tech support.
1. Follow the Cleaning procedure on page 22 to clean the light chamber and optic lenses.
2. Tap
to go to Settings.
3. Tap Calibration.
4. Remove the Meter Check Disk from the foil pouch. DO NOT remove the black cover
from the disk.
5. Insert the Meter Check Disk. Close the lid.
Start
6. Tap
to begin the meter check procedure.
7. If the meter is performing satisfactorily, "Angle Calibration Successful" will be
displayed and the settings will be saved.
If the meter is not performing satisfactorily, "Angle Calibration Unsuccessful. Contact
Tech Support." will be displayed.
8. Tap
to return to the Testing Menu and resume testing.