CAUTION
Avoid continual eye contact with the LED light (Z800BL/Z800WL/Z900BL/Z900WL).
This handpiece is a LED product classified in the "Exempt Group" (no photobiological hazard) in accordance with IEC62471/EN62471.
Use a power source which meets the following requirements (Z800BL/Z800WL/Z900BL/Z900WL).
1. The electricity supply of the power source is below 15W both under normal and single-failure conditions.
2. The power source uses a SELV circuit for electricity supply.
3. The output voltage of the power source is within the range recommended by the manufacturer of this product.
U.S. Federal law restricts this device to sale by or on the order of a licensed physician.