NSK Ti-Max Z800L Manuel D'utilisation page 16

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 56
8-6 Sterilization
Sterilize the product by autoclave sterilization. Remove the bur after each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20min. at 121°C, or 15min. at 132°C, or 3min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required for use.
CAUTION
Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage to the
product from chemical residue on other instruments.
Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt and
sulphur.
Immediately after use, the product should be cleaned, lubricated and sterilized. If blood remains on the external or internal surfaces it can
become clotted and cause rust.
Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
If the sterilizer chamber temperature may exceed 135°C during the dry cycle then delete the dry cycle.
15

Publicité

Table des Matières
loading

Table des Matières