4. Do not introduce air bubbles into the disk. The reagent chambers
will not fill entirely and results will be inaccurate. As soon as a
bubble starts to form, pull back on the plunger to draw the bubble
out of the disk. Begin the filling process again.
5. Wet disks should be dried thoroughly with a lint free wipe. The disk
should be handled by the edges.
6. Disks should be filled and used within 10 minutes. They cannot be filled ahead of time.
For filling and troubleshooting tips go to:
GENERAL OPERATING PROCEDURES
METER When a filled disk is placed in the chamber and the lid is closed, the meter spins at high
speed to distribute the sample to the test wells. Next the meter slows to maximize the pumping
action of the stainless steel mixing beads as the reagents mix with the sample water. Each
reaction is then read at the proper time and wavelength for that reagent system.
The button located in the lower center of the top of the meter turns the instrument on and
off.
The Blue indicator light of the on/off button indicates the status of the instrument.
Steady blue light – the blue LED will remain steady to indicate that the meter is on and ready to
run a test.
Blinking blue light (three blinks/second) – a test is in progress and the disk is spinning. Do not
open the lid when the disk is spinning.
Care should be taken when closing the lid. Do not slam the lid. Wiring between the lid and the
body of the photometer passes through the hinge. The meter will not run with the lid open.
TOUCHSCREEN DISPLAY When the button is
pressed to turn the meter on, the Test Screen will
be displayed.
The display screen is touch-activated. To make a
selection, tap the icon or word on the screen with a
fingertip, fingernail, pencil eraser, or stylus.
¬ Gently wipe smudges from the screen with the
Cloth Wipe (3580-WIPE).
¬ Do not touch the screen with a sharp object.
¬ Do not place objects on the screen that will
scratch or damage it.
¬ Avoid touching the screen with wet fingers.
TESTING
1. Press
and hold until the meter turns on.
2. Tap
. Select a Water Source type. Tap
confirm.
3. Tap
. Select a disk series (found on disk packaging). Tap
NOTE: Disk Series are limited by Water Source Type selection.
4. Remove a disk from the packaging.
5. Use the syringe (1189) to fill the disk with the water sample.
6. Insert the disk. Cover the disk with the Universal Disk Cover (1719). Close the lid.
7. Tap
to start test. Tap
8. The results will be displayed.
www.lamotte.comspin_support
Disk
Type
Power/
Battery
Status
to
to cancel the test. If the test is cancelled discard the disk.
4
Air
Bubble
Water Source
Selection
Source
Treated Water
Disk Series
DW13
16-Apr-20
12:14 PM
Date &
Time
to confirm.
Start
Test
Tap to
Start Test
Test
Settings
History