Transferring results over Bluetooth to a mobile device requires that a mobile app is installed on
the device and an active account for the associated software program. For example, with an
active WaterLink Solutions account and the WaterLink Solutions mobile app on a phone or tablet,
results can be transferred from the meter into the application. Mobile apps for LaMotte software
products are available from iTunes (for iOS devices) and Google Play (for Android devices). Learn
more about LaMotte software products at softwarecenter.lamotte.com.
To transfer results from the WaterLink Spin Touch to a LaMotte software product mobile app:
1. Log into the LaMotte software product mobile app.
2. Search for a customer or Site record. You may need to create one before you can begin
testing.
3. Start a Water Test in the mobile app.
4. Perform a water test from the Spin Touch onboard touchscreen normally. The Spin Touch and
the Bluetooth enabled mobile device will connect automatically.
5. When the connection is available the
meter and device are not connected. Tap the
Whenever the
appears on the touch screen the Spin Touch is capable of connecting to the
BLE Mobile Printer (5-0067). The
not connected.
The Spin Touch cannot connect to both mobile device and the printer at one time. (See Save,
Print and Send Test Results)
USING BLUETOOTH LOW ENERGY The WaterLink Spin Touch will automatically connect to the
nearest Bluetooth enabled phone or tablet or printer. It is not necessary to pair the Spin Touch
to the Bluetooth enabled device. Barriers to wireless signals can reduce the range of wireless
devices. The WaterLink Spin Touch will work best if there are no walls between it and the receiving
devices.
FILLING
When the syringe is placed in the water sample, and the plunger is pulled all the
way up, the syringe will hold more than enough water sample to adequately fill the
disk. Hold the syringe vertically and insert the tip into the fill hole in the disk. Press
the plunger slowly and smoothly to fill the disk.
1. Fill the disk using slow, even pressure. The sample water will fill
the spaces between the baffles in a counterclockwise order. Each
space will fill from the bottom to the top. Sample water should be
added until the sample water in the fourth chamber fills to the top
of the chamber slightly past the embossed fill line. It is OK to fill
slightly past the fill line.
2. Do not overfill the disk. If the disk is overfilled, sample water will
flow out of the overflow hole in the center of the disk. The disk is not
leaking. Dry the disk and run the test.
3. Do not under fill the disk. If the disk is under filled, the reagent
chambers will not fill entirely and results will be inaccurate.
will light on the touch screen. When the is dim, the
to transfer results to the mobile app.
button will be highlighted when connected and dim when
3
Correct Fill
(Fill line)
Overfilled
(Dry disk before placing in meter)
Underfilled