7 Reprocessing
44
7.2

Manual cleaning

IMPORTANT: Manual reprocessing is possible in exceptional cases
if the national/local regulations are followed. The national/local
regulations are to be checked before.
NOTE!  Condition immediately, or at the latest, one hour after
treatment.
NOTE!  Never clean in an ultrasonic bath!
NOTE!  Never immerse in disinfectants!
NOTE!  Use only Midwest Plus Spray.
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
IMPORTANT: The Midwest Handpiece Cleaner is a surfactant.
ü
Wear appropriate protective clothing.
Operating instructions Midwest Stylus Plus / Midwest Tradition
67 27 148  D3673.201.09.01.02