7.3

Manual cleaning and disinfection

IMPORTANT: Manual conditioning is possible in exceptional cases if the
Manual conditioning is possible in exceptional cases
national/regional regulations are followed. The national/regional
regulations are to be checked before.
NOTICE! Condition immediately, or at the latest, one hour after treatment.
Immediately after treatment
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
NOTICE! Only use Midwest® Plus Aerosol Spray.
Using the T1 spray
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Use soft brush
Wear appropriate protective clothing.
Procedure _ Instrument without integrated spray _ without Alpro
All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
1. Brush the instrument under running water (< 38 °C, < 100 °F, at least
drinking water quality) and good lighting (min. 500 lux) and color
rendering index (min. 80 Ra) until no more dirt can be seen, for at
least 10 seconds.
2. Flush the drive channels with spray.
3. Clean and disinfect spray channels with suitable agents and adapters
according to manufacturer's instructions.
4. Blow spray channels with suitable adapters out with 2.5 - 3 bar until
no more moisture displaces, but at least 10 seconds.
5. Conduct thermal disinfection or unwrapped steam sterilization.
6. Maintain mechanical parts manually [ → 15].
7. Maintain the push button chuck manually [ → 16].
8. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
9. Conduct sterilization [ → 17].
7.4

Manual maintenance

7.4.1

Maintenance of mechanical parts

Intervals
● At least every noon and evening
Daily at noon and in the evening
● Prior to each sterilization
● After every thermal disinfection without integrated maintenance
after every thermal disinfection
Required accessories
Required accessories
A
Spray adapter
Spray adapter Midwest E Plus/ E Pro
B
Midwest® Plus Aerosol Spray
B
NOTICE! Only use Midwest® Plus Aerosol Spray.
Using the T1 spray
7 Conditioning
7.3 Manual cleaning and disinfection
15