8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
52
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
8
Maintenance
WARNING
Danger of touching live parts
If the housing is damaged, there is a possibility of touching live parts
inside the unit. If the housing is damaged, the unit must be put and
left out of operation until it has been professionally repaired.
NOTE
Regular inspection
Some countries have legal regulations which require regular safety
inspections of electrical devices or systems by the operator.
Dentsply Sirona would like to draw your attention to the fact that a so-
called "retest" (repeat test) must be carried out for the CEREC AC
acquisition unit at least every three years. In addition, this retest also
must be performed following every repair or retrofit of components
such as the PC, the PC power supply, the CEREC Omnicam camera,
and the camera cable.
NOTE
Annual maintenance performed by trained technical personnel is
recommended.
Care, cleaning, disinfection and sterilization
8.1
NOTE
Use only chemical products recommended by Dentsply Sirona.
8.1.1
Cleaning and disinfection agents
Kerr Corporation
Patterson
Various manufacturers
8.1.2
Monitor screen
Cleaning
NOTE
Never spray the monitor screen with a disinfectant or cleaning spray.
The monitor screen must be wiped off with a soft cloth.
Dentsply Sirona
● CaviCide
● Cavi Wipes
● pdCARE Surface Disinfectant
● pdCARE Wipes
● 60%-90% isopropyl alcohol