7 Operation

7.1 Camera warm-up time

40
Operating Instructions for the Acquisition Unit - With CEREC Omnicam
7
Operation
Camera warm-up time
7.1
When switching on the system, the camera needs to warm up for 15 -
20 minutes. If the coated sapphire glass of the Omnicam is not
sufficiently warm, it may steam up during the acquisition. This
complicates the scan acquisition.
Following use, always position the Omnicam on the heater plate.
You can set the end temperature to which the camera heater warms the
Omnicam mirror sleeve.
1. In the software, navigate to the system menu and click on the
"Configuration" button.
2. Click on the "Devices" button.
3. Click on the "Omnicam" button.
4. Click on the "Camera Heater Settings" button.
5. Use the slider to adjust the temperature.

Adjusting the CEREC Omnicam

7.2
You can adjust the CEREC Omnicam in the device configuration.
1. In the software, navigate to the system menu, and click on the
"Configuration" button.
2. Click on the "Devices" button.
3. Click on the "Omnicam" button.
Accepting settings
➢ Click on the "Ok" button.
Discarding settings
➢ Click on the "Cancel" button.
Dentsply Sirona
66 37 479 D3492
D3492.201.07.28.02    09.2019