Maintenance

9.1

Replacing the control unit fuse

Users can replace faulty control unit fuses themselves. These are located at the rear of the device in the
fuse slot beside the power switch:
Switch off device.
Unplug the power plug.
Open the fuse slot using a screwdriver.
Replace the faulty fuse T 1 AL 250 V AC.
Slide the fuse holder back in and close the fuse slot.
Check the mains voltage shown on the fuse slot.
Plug in the power plug again.
1
1. Fuse slot locking mechanism
2. Display window for voltage setting
3. Fuse slot
9.2

Safety inspections

The essential requirements have been defined and within the risk analysis assessed. The approved results
have been filed in the risk management act with the manufacturer.
The performance of safety inspections on medical devices is required by law in several countries. The safety
inspection is a regular safety check that is compulsory for those operating medical devices. The objective of
this measure is to ensure that device defects and risks to patients, users or third parties are identified in
time.
The STI (Safety Technical Inspection) for the Dispenser DP 30 shall be executed every 2 years by authorized
experts. Results shall be documented.
The service manual, wiring diagrams, and descriptions are available upon request from the manufacturer.
NOUVAG AG offers a safety inspection service for its customers. Addresses can be found in the appendix
of this operation manual under "Service centers". For further information please contact our technical
service department.
Further international service centers are listed on the Nouvag website:
9.3

Information on disposal

When disposing of the device, device parts and accessories, the regulations prescribed by law must be
observed. To ensure environmental protection, old devices can be returned to the dealer or manufacturer.
Contaminated single-use tubing sets are subject to specific disposal requirements. Please observe
prevailing national disposal regulations.
When discarding the device components and accessories, please comply with the issued statutory
regulations. With regard to the preservation of the environment old equipment may be returned to
the distributor or manufacturer.
12
3
2
www.nouvag.com > Service > Service providers
4. Fuse 1
5. Fuse 2
Dispenser DP 30, operating instructions, REF 31678, V52/17
Dispenser DP 30
4
5

Publicité

Table des Matières
loading

Table des Matières