A
A) Press the release key for tubing set bracket (on top of the control unit) to open the pump.
B) The compartment with the integrated tubing bracket opens.
C) Place the tubing set into the tubing bracket provided in such a way that the end of the tubing set
with the spike exits the pump to the rear of the control unit. Check that the tubing is secure.
D) With the tubing set inserted, press the compartment downwards until it clicks into place.
4.
4.
Insert the spike at the end of the tubing set into the irrigation fluid bottle and hang the bottle
onto the stand.
5.
Open the roller clamp on the tubing set as far as it will go.
6.
Open the vent valve at the spike.
7.
Connect the control unit to the power socket.
Ensure that the operating voltage setting corresponds to the local mains voltage.
The container of irrigation fluid may weigh a maximum of 2 kg. Heavier containers than allowed
can cause the device to tip over.
6.3

Device setup

● Place the Dispenser DP 30 and all required accessories and instruments on an even, non-slip surface and
make sure you have good access to all controls.
● Do not allow the operating range of the device (including the cable) to be compromised by limiting
factors.
● The system operation panel and the infiltration liquid bottle must be fully visible at all times.
● It must be explicitly ensured that no objects can fall on the pedal.
● The power plug at the rear of the device must be accessible at all times.