Time; Temperature; Recorder - Consort C731 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Procedure to change or display the time. When switched off, the time
remains resident in the memory for logging and printing purposes.
1.
Press SET until the display shows [SEt tiME] and press CAL to
continue.
2.
The display shows [tiME=on] or [tiME=OFF]. Select the desired mode
with
or
and press CAL to continue or SET to return.
3.
Program the time with
seconds) by pressing CAL each time (or press SET to return).
This procedure permits to adjust the manual or automatic temperature
compensation. For very accurate temperature readings the attached
ATC-probe (Pt1000) can be calibrated in a ±10°C span.
1.
Press SET until the display shows [SEt °C] and press CAL to
continue.
2.
When a Pt1000 temperature probe is connected to the instrument the
display will show the measured temperature e.g. [°C=23.7][Aut]. If
necessary, calibrate your probe with
continue or SET to return. Press RES to reset the temperature
calibration to the original factory calibration.
3.
When no Pt1000 temperature probe is connected to the instrument
the display will show the manual temperature e.g. [°C=23.7][MAn].
Shift to the proper temperature with
continue or SET to return.
Two recorder outputs can be programmed individually to monitor the
measurements of any channel.
1.
Press SET until the display shows [SEt rEC] and press CAL to
continue.
2.
The display shows e.g. [rc1=PH1], [rc1=MV2], [rc1=EC], [rc1=°C],
...etc. Select the desired mode with
continue or SET to return.
3.
This step is skipped unless [rc1=EC] has been chosen. The display
shows the conductivity measuring range & units, e.g. [rc1
0.01MS][Unit]. Select the desired range with
to continue or SET to return.
4.
The display shows e.g. [rc1=6.00][LoW]. Select the desired minimum
level corresponding to a 0 V recorder output with
CAL.
5.
The display shows e.g. [rc1=8.00][HiGH]. Select the desired
maximum level corresponding to a 2 V recorder output with
and press CAL.
6.
The display shows e.g. [rc2=PH1], [rc2=MV2], [rc2=EC], [rc2=°C],
...etc. Select the desired mode with
continue in the same way for recorder-2 or SET to return. The display
shows [SEt rEC].
or
(respectively hour, minutes and
or
or
or
or
or
and press CAL to
and press CAL to
and press CAL to
and press CAL
or
and press
or
and press CAL to

Time

Temperature

Recorder

5

Publicité

Table des Matières
loading

Table des Matières