I-Tech UT2 Notice D'utilisation page 36

Ultrasonothérapie
Masquer les pouces Voir aussi pour UT2:
Table des Matières

Publicité

Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR
The I-TECH UT2 is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such an
environment
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
overvoltage
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supplì input
lines
IEC 61000-4-11
IACER Srl
ALL EQUIPMENT AND SYSTEMS
Test level
IEC 60601
±8kV in contact
±15kV on air
±2kV
for power
supplies lines
±1kV
for input/output
lines
±1kV linea(e) –
linee / line(s) to
line
±2kV line(s) to
earth
<5% U
T
(>95% dip in U
)
T
for 0,5 cycles
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 5s
36
Compliance level
Floors should be
wood, concrete or
ceramic tile. If floors
±8kV in contact
are covered with
±15kV on air
synthetic material,
the
humidity should be
at least 30%.
±2kV
Mains power quality
for power
should be that of a
supplies lines
typical commercial
±1kV
or
for input/output
environment.
lines
±1kV linea(e) –
Mains power quality
linee / line(s) to
should be that of a
line
typical commercial
or
±2kV line(s) to
environment.
earth
<5% U
UT2 con un gruppo
T
(>95% dip in U
)
di continuità (UPS) o
T
for 0,5 cycles
con batterie.
Main power quality
40% U
should be that of a
T
(60% dip in U
)
typical commercial
T
for 5 cycles
or
environment. If the
70% U
user of I-TECH UT2
T
(30% dip in U
)
requires continued
T
for 25 cycles
operation
power
<5% U
interruptions, it is
T
)
recommended that
T
for 5s
I-TECH
powered from an
uninterruptible
Electromagnetic
environment -
guide
relative
hospital
hospital
hospital
during
mains
UT2
be
MNPG165-02

Publicité

Table des Matières
loading

Table des Matières