Télécharger Imprimer la page

SciCan HYDRIM C61 wd G4 Mode D'emploi page 3

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 4
English
Maintenance and troubleshooting
CLEANING THE LCS WATER FILTER
1
Unscrew the cap at
the front of the
LCS adaptor cylinder.
Tube A
Tube B (behind)
LCS system Messages
Replace Water Filter
Replace Air Filter
CF 12 Water
Filter Clogged
CF 18 Water Filter
Pressure Sensor Stuck
CF 26 Air Valve error
CF 30 LCS adaptor
disconnected
2
Remove the filter.
REPLACING THE BACTERIOLOGICAL AIR FILTER
Replace the bacteriological filter
every 500 cycles or if dirty.
To change the filter:
1. Power the unit OFF,
disconnect tube A and remove
the filter from the bracket.
2. Disconnect tube B from the
back of the filter.
What it Means
Water filter in the LCS adaptor
is dirty or clogged.
The bacteriological air filter in the
kickplate is dirty.
Problem with Lumens Cleaning System.
Problem with Lumens Cleaning System.
Problem with air suppy pressure in air line.
Incoming air pressure must be between 1 bar
(15 psi) - 10 bar (145 psi).
Problem with Lumens Cleaning System.
3
Rinse the filter and, if required,
clean with a soft brush or using
compressed air. Check and clean
the O-ring. Remove any debris
inside the housing.
3. Before installing the replacement filter
(01-102119S) check that the arrow mark
on the filter matches the direction of
the arrow on the bracket. (Arrow should
be pointing to the right).
4. Insert filter into bracket and connect
tube B. Re-connect tube A.
5. After replacing the filter, go to the "Reset
LCS Counter" screen in the Setup menu
and reset the counter to 0.
What to Do – User
• Follow instructions above:
"Cleaning the water filter."
• Follow instructions above:
"Replacing the air filter."
• Check LCS adaptor is correctly positioned.
• You selected LCS at program start but
LCS is not installed.
• Replace Water Filter, see
"Cleaning the water filter."
• If cleaning doesn't resolve the problem,
replace the water filter (01-113742S)
• If message persists, call for service.
• Call for service.
• Call for service.
• Check that LCS adaptor is correctly positioned.
4
Return the filter to the
housing and close the
cap tightly.

Publicité

loading