Hach Polymetron 8810 pH Manuel D'utilisation De Base page 3

Analyseur
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 79
Specifications
Specifications are subject to change without notice.
Specification
Ambient temperature
Relative humidity
Operating altitude
Mains power supply
Max. consumption
Overvoltage category
Pollution degree
CE compliance
ETL approved
Korean certification
Compressed air
Rinse water pressure
Reagents
Analog outputs
Alarms
Exchangeable sensors
Temperature compensation
Level control
Pumps
Tubing
Reagent consumption
Titration time
Cycle time
Accuracy
Reproducibility
Details
5 - 40 °C (41 - 104 °F)
10 to 80%
From 0 to 2,000 m. (6,550 ft.) above sea level
110/220/240 VAC, 50/60 Hz, ± 10%
Fuses: 110 to 120V, T630mAL250V; 220V to 240V, T1.25AL250V
110 VA
2 (according to standard EN 61010-1)
2
EN61326-1: EMC Directive
Note: This is a Class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate measures.
EN61010-1: LVD Directive
ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
User Guidance for EMC Class A Equipment
업무용을 위한 EMC 등급 A 장치에 대한
사용자 지침
사용자안내문
A 급 기기 ( 업무용 방송통신기자재 )
이 기기는 업무용 (A 급 ) 전자파적합기기로서 판매자 또는 사용자는 이 점을 주의
하시기 바라며 , 가정외의 지역에서 사용하는 것을 목적으로 합니다.
4 - 7 bar, filtered and dried
0.5 - 6 bar
10 liter storage canister (supplied)
Number: 2; 0 - 20 or 4 - 20 linear programmable
2 x Relay (concentration); 1 x Warning or System alarm
pH + reference electrode
Pt100
Sample; Reagents; Calibration solution; Chemical cleaning
Peristaltic, micro piston, pulse or volumetric pumps for calibration and conditioning
reagents
®
Tygon
; Polyethylene
Application specific
Application specific
Programmable up to 999 minutes
< ± 2 to 4% (application specific)
< ± 2 to 4% (application specific)
English 3

Publicité

Table des Matières
loading

Table des Matières