Power frequency
(50Hz) magnetic
field IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration - electromagnetic immunity
The BM95 is intended for use in the electromagnetic environment specified below. The customer or
the user of the BM95 should assure that it is used in such an environment.
IMMUNITY test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 A/m
IEC 60601 test level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to
2.5 GHz
Not applicable
Compliance level
3 Vrms
3 V/m
217
Not applicable
Electromagnetic environment -
guidance
Portable and mobile RF commu-
nications equipment should be
used no closer to any part of the
BM95 including cables, than the
recommended separation dis-
tance calculated from the equa-
tion applicable to the frequency
of the transmitter.
Recommended separation dis-
tance
d = 1,2 P
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation Dis-
tance in metres (m).