7 Reprocessing
134°C
44
7.5
Sterilizing
ü
274°F
ü
ü
➢
The instrument is clean and disinfected.
The instrument is lubricated.
The instrument can be sterilized in packaging suitable for
sterilization and storage: paper/plastic composite packaging
according to ISO 11607.
Sterilize the instrument in the steam sterilizer with saturated
water vapor.
Overpressure:
Temperature:
Holding time:
2.04 bar (29.59 psi)
134 °C (274 °F)
3 min.
6567460 D3643.201.01.03.02
Instructions for use