7 Reprocessing
36
3.
Pack the instrument in packaging material suitable for
sterilization and storage. e.g. paper/plastic composite
packaging according to ISO 11607.
4.
Perform sterilization  [→ 44].
7.2.2
... with cleaning and disinfection equipment
The instrument can also be cleaned and disinfected in a suitable
piece of cleaning and disinfection equipment.
The cleaning and disinfection equipment used must be approved by
its manufacturer for the cleaning and disinfection of dental
instruments and comply with ISO 15883-1/-2 (e.g., 95°C (203°F)
and 10 min. holding time).
For further details, refer to the instructions for use supplied with the
unit.
ü
The instrument is conditioned with a cleaning and disinfection
device.
Instructions for use
6567460  D3643.201.01.03.02