6 Postprocessing
32
6
Postprocessing
6.1
After each treatment session
NOTE! Condition immediately, or at the latest, one hour after
treatment.
NOTE! Only use W&H Service Oil F1, MD-400.
ü
Wear appropriate protective clothing.
1.
Remove the file with tweezers.
2.
Predisinfect directly at the treatment center [→ 34].
3.
Detach the instrument from the motor.
4.
Transport the instrument to the hygiene room in a suitable
transport container.
5.
Perform automatic reprocessing [→ 35]. Manual reprocessing
[→ 38] is possible in exceptional cases if the national/local
regulations are followed.
Instructions for use
6567460 D3643.201.01.03.02