Sendoline S1-E Notice D'utilisation page 31

Masquer les pouces Voir aussi pour S1-E:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Instructions for use
6567460  D3643.201.01.03.02
evaluation including knowledge of the anatomy of root canals is
important for interpreting the results.
When performing the apex measurement, wear appropriate
insulated gloves. We recommend using a cofferdam for
treatment and putting an insulating sleeve on the contra-angle
handpiece. This will prevent inaccurate measurements caused
by undesired leakage currents. During the measurement, the
instrument must not come into contact with the patient's
mucosa, metallic tooth restorations, or the mucosal electrode.
If the file stops moving, remove the file by gently pulling it
!
toward the coronal end. If this is not possible, switch the motor
to the opposite rotation direction and pull the file out of the
canal.
IMPORTANT: Check the file afterwards for damage or
deformation (untwisting) and replace it if necessary.
5 Operation
29

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

S1-a

Table des Matières