Instructions for use
6567460  D3643.201.01.03.02
7.2

Automated cleaning and disinfection ...

7.2.1
... with a DAC Universal
We recommend using Dentsply Sirona DAC Universal for
automated cleaning, disinfection, and care.
For further details, refer to the instructions for use supplied with the
unit.
ü
The pushbutton chuck is lubricated manually.
ü
The instrument is conditioned using the DAC Universal.
1.
Check whether the instrument is clean after reprocessing
under good lighting (min. 500 lux) and color rendering index
(min. 80 Ra).
2.
If they are still dirty, repeat the process.
Ä
For further reprocessing, the instrument should be dry and
free of residue.
7 Reprocessing
35