Instructions for use
6567460 D3643.201.01.03.02
Only with W&H Assistina
ü
The instrument is clean and disinfected.
1.
Fit the adapter onto the maintenance attachment.
2.
Direct the spray exit nozzle of the adapter downwards.
3.
Press the bur side of the instrument onto the adapter from
below.
4.
Start the Assistina.
5.
Press the instrument onto the adaptor for about 10 seconds.
6.
Remove the instrument from the adapter.
7.
Wipe any spray that comes out with a disinfection cloth.
8.
Close the cover of the Assistina and allow it to run for the
remainder of the maintenance cycle (approx. 25 seconds).
7 Reprocessing
43