Sterilizing - Sendoline S1-E Notice D'utilisation

Masquer les pouces Voir aussi pour S1-E:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operating Instructions S1-E / S1-A
134°C
65 67 460 D3643.201.01.01.02
7.5

Sterilizing

✔ The instrument is clean and disinfected.
Action steps_instruments
274°F
✔ Instrument maintenance is complete.
✔ The instrument can be sterilized in packaging suitable for
sterilization and storage: paper/plastic composite packaging
according to ISO 11607.
➢ Sterilize the instrument in the steam sterilizer with saturated water
vapor.
Unit pressure in bar and psi ww
Overpressure:
Temperature:
Holding time:
2.04 bar (29.59 psi)
134° C (274° F)
3 min
7 Conditioning
43

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

S1-a

Table des Matières