Sendoline S1-E Notice D'utilisation page 44

Masquer les pouces Voir aussi pour S1-E:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

7 Conditioning
42
Only with W&H Assistina
✔ The instrument is clean and disinfected.
1. Fit the adapter onto the maintenance attachment.
2. Direct the spray exit nozzle of the adapter downwards.
3. Press the bur side of the instrument onto the adapter from below.
4. Start the Assistina.
5. Press the instrument onto the adaptor for about 10 seconds.
6. Remove the instrument from the adapter.
7. Wipe any spray that comes out with a disinfection cloth.
8. Close the cover of the Assistina and allow it to run for the remainder
of the maintenance cycle (approx. 25 seconds).
Operating Instructions S1-E / S1-A
65 67 460 D3643.201.01.01.02

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

S1-a

Table des Matières