Operating Instructions S1-E / S1-A
7 Conditioning
7.4.2
Care of the push button chuck
Interval
Interval
● At least once a week
Only with W&H Service Oil F1, MD-400
✔ The instrument is clean and disinfected.
1. Fit the spray cap on the spray can.
2. Hold the instrument firmly.
3. Press the tip of the spray nozzle firmly into the chuck system.
4. Spray into the instrument for about 1 second.
IMPORTANT: Hold the spray can upright.
5. Wipe any spray that comes out with a disinfection cloth.
41
65 67 460 D3643.201.01.01.02