notIce

1. read all of the information in the operation guide and any other literature
in the box before operating the unit.
2. Stay quiet, calm and rest for 5 minutes before blood pressure
measurement.
3. the cuff should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. measuring on same arm for each measurement.
6. please always relax at least 1 or 1.5 minutes between measurements
to allow the blood circulation in your arm to recover. prolonged
over-inflation (cuff pressure exceed 300 mmHg or maintained above
15 mmHg for longer than 3 minutes) of the bladder may cause
z
ecchymoma of your arm.
7.
this Electronic blood pressure monitor s is designed for adults
and should never be used on infants or young children.
Σ
Consult your physician or other health care professionals before
IVD
1
use on older children.
8. Do not use this unit in a moving vehicle, this may result in erroneous
PHTHALATES
measurement.
9. blood pressure measurements determined by this monitor are equivalent
to those obtained by a trained observer using the cuff/stethoscope
auscultation method, within the limits prescribed by the american
national Standard Institute, Electronic or automated sphygmomanometers.
10. Information regarding potential electromagnetic or other interference
between the blood pressure monitor and other devices together
with advice regarding avoidance of such interference please see part
ELECtromagnEtIC CompatIbILItY InformatIon.
11. If Irregular Heartbeat (IHb) brought by common arrhythmias is detected
in the procedure of blood pressure measurement, a signal of (�) will be
displayed. Under this condition, the Electronic blood pressure monitors
can keep function, but the results may not be accurate, it's suggested
that you consult with your physician for accurate assessment.
there are 2 conditions under which the signal of IHb will be displayed:
1) the coefficient of variation (CV) of pulse period >25%.
2) the difference of adjacent pulse period≥0.14s, and the number of such
pulse takes more than 53 percentage of the total number of pulse.
12. please do not use the cuff other than supplied by the manufacturer,
otherwise it may bring biocompatible hazard and might result in
z
measurement error.
13.
the monitor might not meet its performance specifications or cause
safety hazard if stored or used outside the specified temperature
z
Σ
and humidity ranges in specifications.
IVD
1
14.
please do not share the cuff with other infective person to avoid
cross-infection.
PHTHALATES
Σ
15. medical aC adapter which output is DC 6.0V 600ma and complied
IVD
1
PHTHALATES
ENGLISH
24