NOUVAG AG
F F F F
Adjustable torque values:
2, 4, 6, 10, 15, 20, 25, 30, 40 and 50 Nmm

7.6 Adjust "Auto Control" mode

Various modes can be adjusted for controlled root canal treatment. They are activated when the file
reaches the adjusted distance to the Apex. Which ensures that the set minimum distance to the Apex
is reached.
Settings:
The desired mode can be selected with the "Auto control" button → LED of activated mode is
illuminated.
"Auto Slow / Down"
If the set minimum distance is reached during drilling, an acoustic signal is heard and the motor speed
is reduced by half.
If the distance from the tool tip to the Apex is increased again, the motor begins to run at the adjusted
speed again.
"Auto Reverse"
If the set minimum distance is reached, an acoustic signal is heard and the motor switches to reverse
mode.
If the distance from the tool tip from the Apex is reached, the motor returns to normal mode.
"Auto Stop"
If the set minimum distance is reached during drilling, an acoustic signal is heard and the motor stops.
If the distance from the tool tip to the Apex is increased again, the motor begins to run again.

7.7 Hanging up the lip connector

The lip connector act as contact for the Apex-Locator. It is plugged into the lip connector socket and it
is placed in the patient's lower lip.
Check the Apex-Locator function before hanging up the lip connector in the patient's
lower lip!
(see chapter 8. Function control of the Apex- Locator)

7.8 Switching on the motor

The TCM Endo V can be operated with or without foot pedal:
Without footpedal:
The motor can be swithed on and off with the key „Motor"→ If the motor is on, the LED illuminates.
With footpedal:
When using the footpedal, the motor is started and stopped with the step:
Step not pressed:............................Motor switched off
Step pressed down: ........................Motor switched on
The speed of the tool corresponds to the speed which is adjusted on the control unit.