1.1 Use and operation

The TCM Endo V is a microprocessor- controlled endodontic device with Apex Locator.
The Apex Locator measures the distance between file tip and Apex. If a minimum distance is
preselected, the selected auto-control mode is activated, thus ensuring that the minimum distance is
reached.
The speed is held constantly under all circumstances until the maximum adjusted torque limit is
reached. The Automatic Torque Control (ATC) protection mode reduce the risk of a breakage occurs
of the Rimmer. Once the preselected torque limit has been reached, the motor will immediately reverse
for one revolution, then returns to forward direction. Which ensures a fast and effective root canal
preparation.

1.2 Technical data

Voltage: ................................................................................................ 100V∼/115V∼ / 230V∼ / 50-60 Hz
Power supply fuse: ...................................................................................................................... 2 x T1A
Power: ..............................................................................................................................................60 VA
Device protection class: ................................................................................................................ Class II
Application part type: ........................................................................................................................... BF
Rotational speed: .............................................................................................................150 - 2´000 rpm
Precision of measurement: ........................................................................................................ ± 0.2mm
Apex Locator test voltage: .......................................................................................................... 100mV∼
Test voltage frequencies: ........................................................................................... 500Hz and 7.5kHz
Dimensions (H x W x D): ......................................................................................... 115 x 120 x 185 mm
Net weight: ...................................................................................................................................... 1.8 kg
File type: .......................................................................... NiTi-files; shank- Ø 2.35mm (ISO 1797: Typ 1)
1.3

Ambient conditions foroperation

Humidity: .................................. max. 90 %
Temperature: ............................ 10 to 40°C
Pressure: ........................ 800 to 1060 hPa

1.5 Device labels

: Caution: refer to enclosed
documentation
: Application part type BF
1275 : EU conformity symbol
: Certified by the Canadian
Standards Association (CSA) for
Canada and USA
: Device protection class: II
2

1 Product description

1.4
Rel. air humidity: ............................. max. 90%
Temperature: ................................... 0 to 60° C
Air pressure: .......................... 700 to 1060hPa
Ambient conditions for transport
and storage
: Operational earth
: Old electrical and electronic
equipment must be disposed
separately and may not be
included in regular domestic
waste.