1524........................... Control unit TCM Endo V...............................................................................1 unit
1826nou..................... Micro-motor 20 Endo Apex ............................................................................1 unit
1292........................... On / off footpedal ..........................................................................................1 unit
22908......................... Lip connector to Apex Locator .......................................................................1 unit
1809........................... Test plug 0.5mm (± 0.2mm) ..........................................................................1 unit
We are very much concerned with your safety, your team's safety and of course your patient's safety.
Therefore, it is imperative for you to comply with the following instructions:
The TCM Endo V may only be operated by experienced trained personnel!
Please ensure that operating voltage and mains voltage agree!
All sterilisable items must be sterilised before the first use.
(see chapter 9, Cleaning, disinfection and sterilisation).
A performance control of the Apex Locator is to be carried out prior to each
operation (see chapter 8, performance control of the Apex Locator).
The contact bow must constantly touch the file, respectively be folded up as far
as its end position. Otherwise the Apex distance cannot be measured.
The use of third-party products is the responsibility of the operator!
Repairs may only be carried out by authorised NOUVAG service engineers!
Improper use and repair of the device as well as not complying with the
instructions relieve us from any obligations regarding guarantee or other claims!
Caution: Federal law restricts this device to sale by or on the order of a physician
or dentist
Accurate measurement cannot be obtained with the root canal conditions shown below:
No. 31943 35/05

2 Supplied equipment

3 Safety instructions

4 Meausuring Error

Root Canal with a large apical foramen:
Root canal that has an exceptionally large apical foramen due
to a lesion or incomplete development cannot be accurately
measured; the results will show shorter measurement than
the actual length.