7.9 Storage of settings

When switching on the unit, the settings of the last session appear automatically.

7.10 Replacing the file

1)
Fold down contact bow.
2)
Press push-button.
3)
Remove drill.
4)
Release push button.
5)
Insert new file with slight pressure and rotation until it locks into place.
6)
Check whether the file is correctly fixed with a slight pull.
7)
Fold up contact bow until the file lies between the labial bars.
8)
Fold up contact bow up to end position of the file.
9)
Check fit of the contact bow.
RISK OF INJURY!
When operating the device, the contact bow must constantly touch the tool tip,
respectively be folded up as far as its end position. Otherwise, the Apex-distance
cannot be measured and the drilling depth cannot be controlled !
The function control of the Apex-Locator is to be checked prior to each use:
1)
Plug out motor cable from motor socket.
2)
Insert test plug (art. 1809) into motor socket.
3)
Switch on device.
4)
Apex-Locator has to indicate 0.5mm (± 0.2mm)
If the display deviates more than 0.2mm the unit has to be looked-over. Please
contact your supplier in this case.
5)
Plug out test plug from motor socket.
6)
Insert motor cable into motor socket.
7)
Check file fit.
8)
Check contact bow fit.
9)
Hold file to lip connector → Acoustic signal is heard and "LED bar long " is fully illuminated!
If no acoustic signal is heard or if the "LED bar long" is not illuminated, please read chapter 11. Faults
and fault detection.
10

8 Function control Apex-Locator

NOUVAG AG
Klick!
No. 31943 35/05