If buzzer is disabled in the main setup of the balance (see section 13.5) then it remains
disabled. No buzzer settings for check weighing will be shown and the balance is now
ready for check weighing.
If the buzzer is enabled in the main setup of the balance, then next the buzzer setting
options are displayed:
• Press [Up] or [Down] to scroll through the options –
BUZZER
BUZZER
BUZZER
BUZZER
"
OFF"
OFF
OFF
OFF
BUZZER
BUZZER
BUZZER
BUZZER
"
In"
In
In In
BUZZER
BUZZER
BUZZER
BUZZER
"
Out"
Out
Out
Out
• Confirm the Buzzer setting by pressing [Select]
The balance is now ready for check weighing.
• When a weight is now placed on the pan, the display will indicate whether the weight
is below the LOW LIMIT, between the LOW and HIGH LIMITS, or higher than the
HIGH LIMIT, using the capacity tracker, and the buzzer (if set).
• The [Print] key is lit, indicating that print output to a connected device is available.
Print output will be the measured weight as shown on screen.
• To perform check weighing with another sample using the same limits, remove the
sample and place another sample on the top pan.
• To set different check limits, press [Mode] and continue as before
To return to normal weighing, press [Mode] followed by [Select].
© Adam Equipment Company 2015
(Buzzer set to off at all times)
(Buzzer will sound when the weight on the pan is
stable and within the limit)
(Buzzer will sound when the weight on the pan is stable and
outside the set limits)
25
3016612482 Rev1-Aug15