Setting the movement parameters
Select the parameter to be set:
Extension limit, flexion limit, speed, pause at the extension or flexion limit, force or timer; the setting to
change will flash.
Press the
-
or + buttons to modify the setting; the new setting will flash.
To confirm the new setting, press another function button or wait approximately 5 seconds for automatic
confirmation.
Movement parameters can be set either when the machine is stopped or when it is in operation.
At the end of the session, the display reads TIMER 00H00MIN and the unit stops at an average angle use.
Press
-
or + to select a new session time.
Possible values for each parameter:
DAILY COUNTER function: "RESET TIME"
With the splint stopped, simultaneously press the EXTENSION and FLEXION buttons. This function allows you
to see how long the appliance has been operating since it was last reset.
Press the START button twice to reset the counter, or STOP to exit without resetting it.
MANUAL MODE - Start of a session
This function allows you to define and save the maximum limits supported by the patient. This mode is
generally used at the start of a session.
With the Kinetec Performa on STOP, keep the following buttons pressed for at least 3 seconds:
-
to define the extension limit,
+ to define the flexion limit,
and then the movement limit buttons to memorise the actual angle reached.
BY-PASSING MODE - During a session
This function allows you to define and save the maximum limits supported by the patient. This mode is
generally used during a session.
With the Kinetec Performa on RUN, keep the following buttons pressed for at least 3 seconds:
-
to exceed the programmed limit and define a new extension limit,
+ to exceed the programmed limit and define a new flexion limit,
and then the movement limit buttons to memorise the actual angle
Performa
• Extension limit
• Flexion limit
• Speed
• Force
• Extension pause
• Flexion pause
• Timer
Possible values
-3 to 125°
2° to 130°
1 to 7 (from 50° to 220°
1 to 6
0 to 900 seconds (15minutes)
0 to 900 seconds (15minutes)
0 to 24h00
reached.