7. Routine Maintenance

Nimbus 3 and Nimbus 3 Professional Systems
Maintenance
Servicing
Service Period
Nimbus 3 Pump
General Care,
Maintenance and
Inspection
Biofilter
Nimbus 3 and Nimbus 3 Professional Mattresses
General Care
The equipment has been designed to be virtually
maintenance-free between service periods.
ArjoHuntleigh will make available on request service
manuals, component parts lists and other information
necessary for ArjoHuntleigh trained personnel to repair
the system.
It is recommended that the pump is serviced every 12
months by an ArjoHuntleigh authorised service agent.
The service symbol
front panel to indicate that the pump is ready for a
service (Refer to "Service Indicator" on page 13).
Check all electrical connections and the mains power
cord for signs of excessive wear.
Test the Power Fail Alarm before use (Refer to "Testing
the Power Fail Alarm" on page 16).
In the event of the pump being subjected to abnormal
treatment, e.g. immersed in water or dropped, the unit
must be returned to an authorised service centre.
The internal biofilter can be run continuously for two
years before it requires autoclaving or replacement.
The biofilter can only be replaced by a service engineer.
Remove the cover from the mattress.
Inspect the cover for signs of wear or any tears, and
check that all cover fasteners are secure.
Check the security of all internal connections, including:
• Between the cells and the manifold.
• To the CPR/Transport Controls.
• To the Head Section Deflate Control on the
.
Professional
Make sure all cell fasteners are correctly connected to
the mattress base sheet and are not loose or damaged.
25