Arjohuntleigh Nimbus 3 Mode D'emploi page 33

Masquer les pouces Voir aussi pour Nimbus 3:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 205
6. Decontamination
The following processes are recommended, but should be adapted to comply with
the local or national guidelines (Decontamination of Medical Devices) which may
apply within the Healthcare Facility or the country of use. If you are uncertain, you
should seek advice from your local Infection Control Specialist.
The
and
Nimbus 3
decontaminated between patients and at regular intervals while in use; as is good
practice for all reusable medical devices.
Remove the electrical supply to the pump by disconnecting the
mains power cord from the mains power supply before cleaning.
Protective clothing should always be worn when carrying out
decontamination procedures.
Do not use Phenol-based solutions or abrasive compounds or pads
during the decontamination process as these will damage the
surface coating. Do not boil or autoclave the cover.
Avoid immersing electrical parts in water during the cleaning
process. Do not spray cleaning solutions directly onto the pump.
To clean
Chemical Disinfection
Nimbus 3 Professional
WARNING
Clean all exposed surfaces and remove any organic
debris by wiping with a cloth moistened with a simple
(neutral) detergent and water. Dry thoroughly.
To protect the integrity of the cover we recommend a
chlorine-releasing agent, such as sodium hypochlorite,
at a strength of 1,000ppm available chlorine (this may
vary from 250ppm to 10,000ppm depending on local
policy and contamination status).
Wipe all cleaned surfaces with the solution, rinse and
dry thoroughly.
Alcohol based disinfectants (maximum strength 70%)
may be used as an alternative.
Ensure the product is dry before storage.
If an alternative disinfectant is selected from the wide
variety available we recommend that suitability for use
is confirmed with the chemical supplier prior to use.
system should be routinely
Caution
23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Nimbus 3 professional

Table des Matières