Description; Contraindications - Invacare Perfecto2 Série Manuel D'utilisation

Masquer les pouces Voir aussi pour Perfecto2 Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 69

1.3 Description

The Invacare Perfecto
concentrator is used by patients with
2
respiratory disorders who require supplemental oxygen. The device
is not intended to sustain or support life.
The oxygen concentration level of the output gas ranges from 87%
to 95.6%. The oxygen is delivered to the patient through the use
of a nasal cannula.
The Invacare Perfecto
concentrator uses a molecular sieve and
2
pressure swing adsorption methodology to produce the oxygen
gas output. Ambient air enters the device, is filtered and then
compressed. This compressed air is then directed toward one of
two nitrogen adsorbing sieve beds.
Concentrated oxygen exits the opposite end of the active sieve bed
and is directed into an oxygen reservoir where it is delivered to the
patient.
The Invacare Perfecto
concentrator is capable of operation by the
2
patient in a home environment or in an institutional environment.
Device operates at a nominal 230 VAC/50 Hertz supply.
Both concentrator models (IRC5PO2AW and
IRC5PO2VAW) have identical operating parameters with the
exception of the product weight and power consumption.
Refer to typical product parameters.
Service information will be available upon request to qualified
technical personnel only.

1.4 Contraindications

Invacare is not aware of any contraindications for the Invacare
Perfecto
Concentrator Series.
2
1163146-H~07
General
7

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Perfecto2 irc5po2awPerfecto2 irc5po2vaw

Table des Matières