Technical data

Dräger Testor 2100
Ambient conditions
During operation
Temperature
Air pressure
Rel. humidity
During storage
Temperature
Air pressure
Rel. humidity
Housing
Material
Dimensions (D x H x W)300 x 335 x 515 mm
Color
Weight
Compressed air supply
Compressed air connectionDräger PA plug-in nipple
Connection for lung demand valve
Inlet
Outlet
Test head
Medium pressure manometer
Measuring range
Scale division
Class
Pressure relief valve
Low pressure manometer (depending on version)
Measuring range
Scale division
Class
Measuring range
Scale division
Class
Timer
Preselectable
measurement period
After end of time
Battery
32
o
+10 to +40
C
850 to 1400 mbar
max. 70 %
o
–30 to +60
C
850 to 1400 mbar
up to 100 %
ABS plastic
blue / light gray
approx. 4,9 kg
6 bar (4 bar to 9 bar),
breathing air quality
corresponding to EN 12 021
Dräger PA plug-in coupling
round thread Rd 40x1/7"
(internal thread)
according to EN 148-1
rubber, inflatable and
detachable
0 bar to 16 bar
0.5 bar
1.6
integrated
–12 / 0 / 28 mbar
1 mbar
1.6
–30 / 0 / 30 mbar
0.5 mbar
1.0
LCD display
1 sec. to 99 min.
acoustic signal
coin cell 1.5 V - LR 44
Dräger Testor 3100
Ambient conditions
During operation
Temperature
Air pressure
Rel. humidity
During storage
Temperature
Air pressure
Rel. humidity
Measured value displays of the sensors
Low pressure
value
Medium pressure
value
High pressure
value
Operating characteristic values
Dimensions
Depth x height x width 300 x 335 x 515 mm
Weight
Operating voltage
o
+10 to +45
C
850 to 1500 hPa
30 to 70 %
o
–10 to +60
C
850 to 1500 hPa
30 to 70 %
± 30 hPa/ ±1 % from the final
2.5 MPa/ ±1 % from the final
35 MPa/ ±1 % from the final
approx. 5 kg
5 V DC via USB

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Testor 3100

Table des Matières