7.
After the denture is secured, apply Lucitone Digital Fuse Step 2 - 3D Denture Bonding Resin
to the repair area and tack cure for 10 seconds using the UV Tack-Cure Light. Repeat this
process as needed to achieve the desired repair thickness (Figures 38A, B, C & D)
Figure 38A
8.
Apply Lucitone Digital Fuse
to the repair area. (Figure 39)
9.
Place the appliance repair side up in the inLab Speedcure
Processing Unit and select "Lucitone Digital" and then
"Repair." The program will run for 10 minutes followed by a
3-minute cool down. (Figure 40)
Note: Only cure one repair denture at a time. Only cure one
side of the arch.
10.
Finish and polish the denture using conventional techniques. (Figures 41A, B, C & D)
Figure 41A
CAUTION: Do not steam clean Lucitone Digital Print appliances.