Attachments

Attachments recommended by IKA must be connected to the
agitation table to ensure safe and secure operation. in addition ,
the usable weight and speed ranges shown in the diagrams (hat-
ched areas) must not be exceeded.
AS 260.1 Universal attachment
Universal, steplessly adjustable clamping cylinders make it posi-
ble to adjust to any container shape.
The universal attachment consists of:
AS 1.60 basic suspension
AS 1.5 fastening screw
AS 1.61 clamping cylinder
ATTENTION: The base suspension of the universal attachment
is secured in place by four lateral clamping screwson the agitati-
on table.
AS 260.2 Holding bracket adapter
With additional holding brackets (AS2.1, AS2.2, AS2.3, AS2.4
and AS2.5) the holding bracket adapter is suitable foe working
with round flasks, measuring flasks and Erlenmeyer flasks.
Recommended / maximum fitting
AS 2.1 Holding bracket
AS 2.2 Holding bracket
AS 2.3 Holding bracket
AS 2.4 Holding bracket
AS 2.5 Holding bracket
ATTENTION! The holding bracket adapter is secured in place
by four lateral clamping screws on the agitation table.
AS 260.3 Dish attachment
For gentle agitation at low speed, for example for Petri dishes or
culture flasks. The dish attachments is equipped with an anti-slip
sheet that prevents the dish from moving around during the agi-
tation process.0
28
ATTENTION! The dish attachment is secured in place by four
lateral clamping screws on the agitation table.
AS 260.5 Separating funnel attachment
For intensive agitation of sparating funnels.
Max. fitting:
6 x 50 ml Separating funnel
5 x 100 ml Separating funnel
3 x 250 ml Separating funnel
1x
Dimension WxHxD
3 x 500 ml Separating funnel
8x
in mm
4x
425 x 135 x 334
ATTENTION! The separating funnel attachment is secured in
place by four lateral clamping screws on the agitation table.

Supported weight (load)

In order to ensure safe and secure operation, the agitator must
only be operated within the range of usable weight and speed
(see diagram - hatched area). Please make certain that the hol-
ding surface on which the agitator is placed is clean and level.
12 / 12
Dimension WxHxD
Otherwise the working range shown in the diagram (the hatched
8 / 9
in mm
area) cannot be used.
5 / 8
330 x 24 x 334
If the agitator is screwed together with an adjacent piece of
4 / 4
equipment (i.e. if it is integrated into a system) it should be noted
2 / 4
during operation that dynamic forces may arise as a result of
unfavorable loading circumstances and the position of the center
of gravity. This may cause the adjacent equipment to vibrate or
may cause the table to vibrate uncontrollably. If you notice that
the device is running unevenly, the speed must in any case be
reduced until the running irregularities no longer occur.
Make certain that individual agitation vessels are placed in the
middle of the agitation table and multiple agitation vessels are
Dimension WxHxD
in mm
410 x 33 x 370
Dimension WxHxD
in mm
228 x 95 x 234
HS/KS260 0305