Specifications

Size
Weight
Electrical Requirements
3655DX & 3655TX
3655IX
3655KX & 3655PX
Sound Level
Power Consumption
3655DX & 3655TX
3655IX, 3655KX & 3655PX
Maximum Compressor Pressure
Compressor Free Air Flow
Operating Temperature Range
Operating Humidity
Storage/Transport Temperature Range
Storage/Transport Humidity
Supplied Nebulizer Capacity
Supplied Nebulizer Nebulization Rate
Supplied Nebulizer MMAD
*Conditions may vary based on altitude above sea level,
barometric pressure, and temperature.
**Value determined with minimum flow rate through
the nebulizer (20°C, sea level).
There are no hazardous materials used in the construc-
tion of this unit, therefore, there are no known risks
associated with disposal of the equipment.
This product (models 3655DX & 3655TX) complies
with the following electromagnetic compatibility stan-
dard: IEC 60601-1-2:2nd Edition
1
A-3655X
7.5" W x 4.0" H x 6.8" D
(19.1 cm x 10.2 cm x 17.3 cm)
4.1 lbs. (1.86 kg)
2.5 amps, 115 VAC (±10%) 60 Hz
1.4 amps, 230 VAC (±10%) 50 Hz
1.4 amps, 230 VAC (±10%) 60 Hz
56 dBA
140 watts maximum
160 watts maximum
35 psig or greater (241 kPa)*
9 lpm or greater*
+40° to +104°F (+5 to +40°C)
up to 95% non-condensing
-40° to +158°F (-40° to +70°C)
up to 95% non-condensing
6 ml (cc)
0.15 ml/min or greater*
5 microns or less**
3655DX & 3655tX
equipment Classifications:
Protection against electric shock: Class II
Degree of protection against electric shock:
Type BF
Equipment not suitable for use in the presence of
flammable anaesthetic mixture with air or with
oxygen or nitrous oxide.
Mode of operation: Continuous operation