1.2 Specifications Titrator TitroLine
CE Zeichen:
Country of origin:
The following solvents/titration reagents are allowed to be used:
− All common titration solutions.
− As reagent water and all non-aggressive non-organic and organic fluids are allowed. If using combustible
fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the chemical industry.
− For fluids with higher viscosity (≥ 5 mm
speed can be adjusted.
− Fluids with viscosity over 10mm
Measuring input:
pH
pH
U [mV]
mV
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
TitroLine® 5000 titrator to "warm up".
Measurement input:
Measurement
T [°C]
* The measurement uncertainty of the sensor probe has to be taken into account as well
Display:
Calibration:
Inputs:
Input pH/mV:
Input Pt 1000/NTC 30:
Power supply:
As per the low voltage directive 2014/35/EU;
test basis EN 61 010-1: 2011-07 for laboratory devices
As per the EMC directive 2014/30/EU; test basis EN 61 326 Part 1: 2012
As per the EMC directive 2011/65/EU; test basis EN 50 581: 2013-02
FCC part 15B and ICES 003
Germany, Made in Germany
2
/s), lower boiling point or affinity to outgas, the filling and dosage
2
/s cannot be dosed.
pH/mV-input with 12 bit transducer for high-precision readings.
Electrode socket according to DIN 19 262 or
additional with BNC socket insert (Z 860)
Reference electrode 1 x 4 mm socket
Measurement range
- 3.0 ... 17.00
- 1900 ... 1900
Temperature probe - connector for a Pt 1000 resistance thermometer and NTC 30
Connection: 2 x 4 mm - sockets and 1 x 2 mm socket.
Display resolution
range
- 30 ... 115
3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Automatically with up to three buffer solutions, sequence during calibration optional,
freely definable buffers can be input.
Default buffer solutions according to DIN 19 266 and NBS, or technical buffers:
pH = 1.00; pH=4.00; pH=4.01; pH=6.87; pH= 7.00; pH= 9.18; pH=10.00;
pH/mV-input with electrode socket according DIN 19 262/or BNC
Temperature sensor probe (Connection sockets: 2 x 4 mm
power supply 100 – 240 V; 50 – 60 Hz, power input: 30 VA
Use the Power supply TZ 1853, Type No.: FW 7362M/12 only!
®
5000
Display
Measurement
resolution
accuracy* without
0,05 ± 1 Digit
0.01
1
1,0
Measurement
accuracy* without
Pt 1000 sensor
0.5 K ± 1 Digit
0.1
(Release: 30. August 2018)
Input
resistance
[Ω]
sensor probe
> 5
± 1 Digit
> 5
Measurement
accuracy* without
NTC 30 sensor
0.5 K ± 1 Digit
12
10
12
·
10