Cleaning - arjo Minuet 2 Notice D'utilisation

Masquer les pouces Voir aussi pour Minuet 2:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

7. Cleaning

Disconnect the bed from the electricity supply before cleaning.
The bed should be cleaned and disinfected before it is used by a new
patient.
These instructions also apply to accessories but not to lifting straps
and handles, mattresses or safety side pads.
Cleaning
Wearing suitable protective clothing, clean all surfaces with a
disposable cloth soaked in a neutral detergent and hand hot
water.
Start by cleaning the upper parts of the bed and clean along all
horizontal surfaces. Work methodically towards the lower
sections of the bed and clean the castors last.
Take extra care with areas that may trap dust or dirt.
Rinse with clean water and dry with disposable paper towels.
Allow the cleaned parts to dry completely before returning the
equipment to use.
Disinfecting
After cleaning the bed, wipe all surfaces with sodium
dichloroisocyanurate (NaDCC) at a concentration of 1,000
parts per million (0.1%) of available chlorine.
In the case of pooling body fluids, e.g. blood, increase the
concentration of NaDCC to 10,000 parts per million (1%) of
available chlorine.
Do not use abrasive compounds or pads as these may damage the
finish.
Do not use phenol-based disinfectant solutions.
Avoid splashing the actuators, control box and control handset.
WARNING
Caution
52

Publicité

Chapitres

Table des Matières
loading

Table des Matières