13. Disposal

Electronic devices are recyclable waste and must not be placed in household waste. At the end of its ser-
vice life, dispose of the product according to the relevant statutory regulations. Remove the inserted battery
and dispose of it separately from the product.

Disposal of used batteries/rechargeable batteries!

You are required by law to return all used batteries. They must not be placed in household waste.
Contaminated batteries/rechargeable batteries are labelled with symbols to indicate that disposal in do-
mestic waste is forbidden. The designations for the heavy metals involved are: Cd=cadmium, Hg=mercury,
Pb=lead (the marking can be seen on the battery, e.g., underneath the refuse bin symbol shown on the left).
Used batteries can be returned to local collection points, our stores or battery retailers.
That way you fulfil your statutory obligations and contribute to the protection of the environment!

14. Troubleshooting

In purchasing this measuring device, you have acquired a product which has been designed to the state of
the art and is operationally reliable. However, problems and malfunctions may still occur. This section tells
you how to troubleshoot common issues:
Error
The device does not work.
The reading does not change.
The display shows "- - - -"
Wrong pH readings are displayed
The measuring device cannot be
calibrated in the pH measuring
function.
The pH electrode has dried up.
Any repair work other than that described above must be carried out by an authorised technician.
If you have questions about the measuring instrument, please contact our technical support team.
Possible cause
Is the battery empty?
The HOLD function is active.
The measuring range has been
exceeded.
No temperature compensation was
performed.
No calibration was performed.
The pH electrode is consumed.
The reading is out of tolerance for
automatic calibration.
The pH electrode is consumed.
The moisture container is empty or
not tight.
Solution
Check the status. Replace the bat-
tery if required.
Press the "HOLD" key.
The measuring device is not
equipped for this range.
Perform the temperature compensa-
tion before each measurement.
Perform a 2-point calibration before
each measurement.
Replace the pH electrode with a
new one.
Adjust the reading on the lateral
matching controllers.
Replace the pH electrode with a
new one.
Try a regeneration of the electrode
or replace the pH electrode with a
new one.