Note:
The heating function can only be started when the cover is closed. When the cover
is opened, the shaker automatically switches off the shaking and heating functions.

Operating modes

You can choose between operating modes
Appliance does not start up again following power outage
Safety limit value for speed and temperature adjustable.
Appliance starts up again following power outage
Safety limit value for speed and temperature adjustable.
Appliance starts up again following power outage
Safety limit values cannot be changed.

Setting the operating mode

Switch on appliance and simultaneously hold down the Time Start/Stop key
and the shift key
.
The display indicates either
(factory default setting A).
Switching the appliance off and on switches to the next operating mode .
Following the selection of the operating mode, the corresponding letter
appears on the display for 5 seconds, the appliance is ready for operation.
and
or
depending on the operating mode set

External temperature sensor

If external temperature sensor PT1000.60 is connected to the internal slide-on
receptacle, the temperature can be measured at any place in the medium.
Locate the external temperature sensor and
the connecting cable so that the shaking
motion is unimpeded and no vessels are
damaged or tipped over.
Calibration - temperature
The appliance is calibrated at the factory.
This function allows the temperature to be calibrated to a desired value.
This may be necessary, for example, if special sample containers, their
arrangement or external influences (e.g. sunshine) cause the measuring value for
the temperature to be incorrect.
(Connector for PT1000.60)
The temperature control of the heating works
automatically with this display or measuring value. This
is indicated by the lighting up of the LED next to the
temperature sensor symbol.
19