Reset; Appliance Variant Ks 4000 Ic Control - IKA KS 4000 i control Mode D'emploi

Masquer les pouces Voir aussi pour KS 4000 i control:
Table des Matières

Publicité

Calibration without inserted sensor:
- Fill Erlenmeyer flask (250ml) with water to the 100ml mark
- Place the Erlenmeyer flask in the centre
- Immerse the sensor of the external measuring device in the water
- Set the target temperature
- Close the cover and start the Temp function
- Wait until the temperature in the incubation cover has stabilised.
- While holding down the shift key
display appears; keep holding down the shift key
- Set the temperature value read off the external measuring device using the
Temp up  down  keys
- Calibration is complete when you release the shift key.
Calibration with inserted sensor:
Calibration with inserted sensor takes place in the same way as described above.

Reset

To restore the factory settings, hold down the Mot  key and the Temp  key
and press the main switch I/O.
Factory setting:
Operating mode A
Upper speed limit 500 rpm
Safety temperature limit 90 °C
Temperature calibration.
, briefly press the Temp On/Off key
.

Appliance variant KS 4000 ic control

There is a cooler built into this appliance variant.
By using an external cooling unit, the working temperature can be lowered in
relation to the room temperature (depending on the supply temperature).
The cooler is connected to an external cooling unit (e.g. IKA KV 600) via the
plug connection at the back of the appliance. The inlet and outlet connections
, the
are labelled accordingly at the back of the appliance. To connect the cooling
pipes there are two hose connection pieces included. They can be connected to
a hose with a 10mm interior diameter. The connectors are unlocked by coaxially
positioning the unlocking lever and pressing slightly in the direction of the arrow.
By positioning and light pressure in the direction of the plug, the connectors are
locked/connected to the inlet/outlet plug connections on the housing.
Drain hose
Unlocking lever
Plug connection
cooler inlet
Plug connection
cooler outlet
20

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ks 4000 ic control

Table des Matières