Install The Battery - Hach 2100Q Manuel D'utilisation Simplifié

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
Section 3 Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the
tasks described in this section of the document.

3.1 Install the battery

W A R N I N G
Explosion hazard. An expired battery can cause hydrogen
gas buildup inside the instrument. Replace the battery before
it expires. Do not store the instrument for long periods with a
battery installed.
W A R N I N G
Potential fire hazard. Use only alkaline or nickel metal hydride batteries
(NiMH) in the meter. Other battery types or incorrect installation can
cause a fire. Never mix battery types in the meter.
N O T I C E
The battery compartment is not waterproof. If the battery compartment
becomes wet, remove and dry the batteries and dry the interior of the
compartment. Check the battery contacts for corrosion and clean them
if necessary.
N O T I C E
When using nickel metal hydride (NiMH) batteries, the battery icon will
not indicate a full charge after freshly charged batteries have been
inserted (NiMH batteries are 1.2 V versus 1.5 V for alkaline batteries).
Even though the icon does not indicate complete charge, 2300 mAH
NiMH batteries will achieve 90% of instrument operation lifetime
(before recharge) versus new alkaline batteries.
N O T I C E
To avoid potential damage to the meter from battery leakage, remove
the meter batteries prior to extended periods of non-use.
The meter can be powered with AA alkaline or rechargeable NiMH
batteries. To conserve battery life, the meter will power off after
10 minutes of inactivity, the backlight powers off after 30 seconds. This
time can be changed in the Power Management menu.
Note: Rechargeable batteries will only be recharged with the USB/power module.
Refer to the module documentation for further information.
For battery installation refer to
Figure
1. Remove the battery cover.
2. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct orientation.
3. Replace the battery cover.
Figure 3 Battery installation
3.
English 7

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

2100qis

Table des Matières