Hach 2100Q Manuel D'utilisation Simplifié page 11

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
compartment so the diamond or orientation mark aligns with the raised orientation
mark in front of the cell compartment.
1. Push the
2. Insert the 20 NTU
CALIBRATION key
StablCal Standard
to enter the
and close the lid.
Calibration mode.
Note: The standard to
be inserted is bordered.
Follow the
instructions on the
display.
Note: Gently invert
each standard before
inserting the standard.
4. Repeat Step 2
and 3 with the
100 NTU and
3. Push Read. The
800 NTU StablCal
display shows
Standard.
Stabilizing and then
Note: Push Done to
shows the result.
complete a 2 point
calibration.
7.2 Reading modes
1. Push the UP or DOWN key to enter the Reading Options menu.
2. Select Reading Mode to select one of the following options:
5. Push Done to
review the
calibration details.
Option
Description
Normal
The normal mode reads and averages three
(Default
readings. The result is shown after the reading.
setting)
6. Push Store to
save the results.
After a calibration is
complete, the meter
automatically goes
into the Verify Cal
mode. Additional
information is
available on the
manufacturer's
website.
English 11

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

2100qis

Table des Matières